Specifications |
CAPNOGARD®
Capnograph
PRINCIPLE OF OPERATION Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.
INITIALIZATION TIME Capnogram within 15 sec., full specifications within 60 sec.
CALIBRATION Routine calibration is not required. 15 sec. adapter zero performed when changing to different style of adapter.
ACCURACY VERIFICATION Accuracy verifier provided on sensor cable.
CAPNOSTAT® MAINSTREAM CO2 SENSOR
| Size: |
1.30 * 1.67 * 0.85 inches (3.30 * 4.24 * 2.16 cm) |
| Weight: |
0.64 oz. (18 grams) - cable excluded |
| Shock Resistance: |
Withstands repeated 6 foot drops |
AIRWAY ADAPTERS Adult/pediatric reusable or single patient use (added dead space <5cc), neonatal reusable (added dead space <.5cc), and sampling adapter for non-intubated patients.
SAMPLING FLOW RATE 180 ml/min (for non-intubated patients).
CO2 (CARBON DIOXIDE)
| Compensations: |
N2O & O2 (selectable), BP (automatic) |
| Capnogram: |
Selectable sweep speed (12.5 or 25 mm/sec) |
| Range: |
0 - 100 mmHg |
| Accuracy: |
±2 mmHg (for 0 - 40 mmHg) ±5% of reading (for 41 - 70 mmHg) ±8% of reading (for 71 - 100 mmHg) |
| Resolution: |
1 mmHg |
| Response Time: |
60 ms |
RESPIRATORY RATE
| Accuracy: |
±1 br/min |
| Range: |
0-150 br/min |
| Resolution: |
1 br/min
|
General
ALERTS
| Limits: |
Adjustable limits for ETCO2, SpO2, Respiratory and Pulse Rates |
| Audio: |
Adjustable volume, 2 min. silence or OFF (LED indicators) |
| Visual: |
On-screen & red "Alert Bar" |
DISPLAY
| Type: |
Dot matrix, Cold Cathode Display (CCD) |
| Size: |
5" W * 1.5" H (12.7 * 3.08 cm) |
GRAPHIC TREND/HISTOGRAM
| Memory: |
24 hr., battery backed |
| Format: |
On-screen or printed, in 30 min., 2, 8, or 12 hr. segments
|
COMMUNICATION OUTPUT
| Digital: |
RS232 (printer, NovaCARD, computer interface) |
| Analog: |
Optional module, 0-1VDC (numerics & waveforms) |
PHYSICAL
| Size: |
3.3" H * 9.0" W * 8.0" D (8.38 * 22.86 * 20.32 cm) |
| Weight: |
8 lb (3.63 kg |
ELECTRICAL
| Power requirements: |
100-120/200-240 VAC, 50-60 Hz, 30 VA |
| Battery: |
Sealed lead acid gel cell, 2 hr. life, 12 hr. recharge |
ENVIRONMENTAL
| Operating temperature: |
50-104° F (10-40° C) |
| Operating humidity: |
0-90% relative (non-condensing)
|
|